IDA (Interdigitated Array) Electrode - 2 µm

More fine IDA (Pt/Au) Electrode has been developed. Its 2 µm pitch working electrodes significantly increase redox cycle and allow researchers to perform high-ultra-sensitive measurement.
IDA(Interdigitated Array) Electrode Au 2 um

Technical data concerning IDA Electrode 2 µm

A CV curve obtained by 2 µm IDA Electrode Au.
 

CV by Single mode

CV (Single) Using Au 2 um

  • Sample: 1 mM ferrocene methanol (0.5 M NaCl solution)
  • Mode: Single
  • Sweep Rate: 10 mV/s

CV by Dual mode

CV (Dual) Using Au 2 um

  • Sample: 1 mM ferrocene methanol (0.5 M NaCl solution)
  • Mode: Dual
  • Sweep Rate at G-Electrodes: 10 mV/s
  • Applied Voltage to C-Electrodes: -0.05 V
    Its collection efficiency was proved to be 97% by our calculation from limiting current of Collector electrodes and Generator electrodes.

Capture Efficiency

Compared with IDA Au 10 µm and 3 µm, the capture efficiency of the 2 µm is higher, 93% for 10 µm and 95% for 3 µm.
For 2 µm electrode, the limiting current of ferrocene methanol by CV Dual mode was 11 times as much as of the peak current by Single mode, while this ratio for 10 µm and 3 µm electrode were about 4 and 7.5, respectively. This means large increase of the redox cycle.


 

New Product

new
New width in the product line, and you could have 2 µm IDA with or without passivation membrane.

Catalog No. Description Width (µm) Interval (µm) Length (mm) Number of feet (pairs)
012257 IDA electrode (Au) 2 2 2 65
012258 IDA electrode (Pt) 2 2 2 65
012261 IDA electrode (Au)
without passivation membrane
2 2 2 65
012264 IDA electrode (Pt)
without passivation membrane
2 2 2 65



 

Accessories

Catalog No. Description
011066 Cable kit for IDA electrode
011464 Ag/AgCl Ink for reference electrode (2.0 mL)