<<Home

Conductivity Electrode

 

The Principle of Conductivity measurement

When the electrical resistivity of a conductive polymer sheet is high(R>10KΩ), its conductivity can be obtained by Two electrodes measurement. On the other hand, Four electrodes measurement with guard electrodes is applied when that of conductive polymer sheet is low (R<10KΩ). This method is effective to eliminate fault resistivity of the lead wire or its contact point with the sample. Actual layout of four electrodes measurement is shown in Figure1. Constant current [I] from the constant-current power supply [S] runs through the sample via the two electrodes for current [C1],[C2]. Its potential difference [UP1P2] between the other two electrodes for potetial-difference [P1], [P2] is measured by POT - the potential difference detector with high input impedance. In this measurement, the impedance [ZP1P2] of [UP1P2]=[P1P2], potential difference [UP1P2] and constant current [I] are linked by the relation:UP1P2 = IZP1P2. Therefore, the conductivity can be calculated from the formula mentioned below:

Formula_for_Condyctivity
[Fig.1]
noaround,nolink,zoom,Fig.1
S:constant-current power supply,
POT:conductivity meter Model1116,
RS:standard adjustable resistor,
C1,C2:electrodes for current,
P1,P2:electrodes for potential difference


 

Conductivity Meters

Model_1116SLD
 

  • Model116SLD Manual (4.88 MB) Model116SLD Model116SLD


     
Description Model 1116 Conductivity Meter for Solution Model 1116SLD Conductivity Meter for Solid
Catalog No. 010437 010438
Range of measurement 2,000 / 200 / 20 / 2 mS (1KHz) 200mΩ / 200 KΩ / 2 / 20 / 200 MΩ (

Our Conductivity Meters have high tolerance for external noise since op-amp and phase sensitive detector are equipped. Not only that the obtained value is shown their front panel digital display, but also that users can run serial measurement by rear output panel. Model 1116SLD is specialized for solid sample because solution is decomposed by its DC current whereas Model 1116 is capable of measuring liquid sample by AC method.


 

4-probes electrode for conductivity measurement

4-probes electrode
Catalog No. Description
010439 4-probes electrode for conductivity measurement


Model 1116 SLD equips a cell cable with alligator clip as a standard accessory, however we recommend to utilize our 4-probes electrode for conductivity measurement in order to obtain accurate value. The four electrodes steadily contact with the sample by 50gf of needle pressure and rhodium plating strengthens their surface. There is no need to replace entire electrodes if one of them became dysfunctional - respective four electrodes can be replaced seperately.


 

4-probes electrode--contact part Replacable Electrodes
Measurement Example--Au Measurement Example--Au2
Measurement Example--GC Measurement Example--ITO

Example of Measuring Accuracy of Model1116 SLD
This table shows relation amang range, measuring current and measuring accuracy.

Range(Ω) 200m 2 20 200 2K 20K 200K 2M 20M 200M
Measuring Current 100mA 10mA 10mA 10mA 1mA 100μA 10μA 1μA 100nA 10nA
Measuring Accuracy 1% 10% 0.5% 0.5% 0.5% 0.5% 0.5% 1% 2% 2%








 

Conductivity electrode

Conductivity Electrode





 

Catalog No. Description
011315 Conductivity electrode







Platinum terminals are deposited on a fused quartz substrate as current supplying electrodes and potential difference probing electrodes. The distance between electrodes for potential difference are adjustable from 40 µm to 250 µm by changing the connect pin.

<<Home                                  <<Back to ALS Main page