Plate Material Evaluating Cell
This quite handy cell is developed in order to
evaluate plate material such as metal,
semi-conducting plate, etc. In evaluation, a sample
plate is sandwiched between cell blocks.
Catalog No. | Description | Quantity |
011951 | Plate Material Evaluating Cell | 1 |
Contents | ||
Teflon Cell [Body] | 1 | |
Teflon Cell [Base] | 1 | |
O-ring | 1 | |
Screw 20 mm | 2 | |
Purging tube 1 m | 1 | |
002222 | Pt counter electrode for VC-2 cell | 1 |
[Optional Products] Reference Electrode
Users can select a reference electrode separately for this cell from :
Catalog No. | Description |
012167 | RE-1B Silver-Silver chloride reference electrode |
012171 | RE-7 Non Aqueous reference electrode |
002058 | RE-1C Silver/Silver Chloride Reference Electrode (Saturated KCl) |
002056 | RE-2B Calomel Reference Electrode |
002057 | RE-2C Reference Electrode |
011503 | RE-6H Reference Electrode Kit available in alkaline solution |
How to Set Up
1. Place a plate sample on a lower side teflon
block of the cell.
Applicable plate dimensions : W: 10-20 mm, T: ~3 mm
2. Put another teflon block on the sample, and drive
two screws evenly to fix the plate. Liquid contact
area depends on inner diameter of the O-ring.
Deliver approximately 1 ml of electrolyte after
fixing the plate.
3. Attach a Pt counter electrode and a reference
electrode to the cell by setting them onto a cell
cap. We would recommend you to twist up the Pt part
appropriately if the counter electrode is longer.
<<Home ALS Main Page